What Is Measurement Uncertainty? A Simple Guide
What Is Measurement Uncertainty?
Measurement uncertainty is a number that describes how much doubt exists about a measurement result. No measurement is perfect, so a complete result includes both the value and its uncertainty, for example 12.34 ± 0.05 cm. This says the true length is very likely somewhere between 12.29 and 12.39 cm.
The internationally accepted method is described in the Guide to the Expression of Uncertainty in Measurement (GUM). See the measurement uncertainty entry on Wikipedia for background.
Uncertainty vs Error
Error is the difference between a measured value and the true value, which you usually can’t know exactly. Uncertainty is an estimate of how large that error could reasonably be. You correct known errors; you report the remaining uncertainty.

Type A and Type B Uncertainty
| Type | How It’s Evaluated | Examples |
|---|---|---|
| Type A | Statistically, from repeated measurements | Scatter in 10 repeated readings |
| Type B | From other information | Calibration certificate, instrument resolution, manufacturer specs |
For Type A, the standard uncertainty of an average is the standard deviation divided by the square root of the number of readings: u = s ÷ √n. For a Type B source given as a range of ±a with no preferred value inside it, such as a digital display’s resolution, the standard uncertainty is usually a ÷ √3.
Combining Uncertainties
Independent uncertainty sources are combined by the root-sum-square method:

uc = √(u1² + u2² + u3² + …)
The result is then usually multiplied by a coverage factor k = 2 to give the expanded uncertainty U, which corresponds to roughly 95% confidence.
Worked Example
A student measures a block’s length five times with a digital caliper (resolution 0.01 mm): 25.42, 25.44, 25.43, 25.45, 25.41 mm.
- Average: 25.430 mm.
- Type A: standard deviation ≈ 0.0158 mm; uA = 0.0158 ÷ √5 ≈ 0.0071 mm.
- Type B (resolution): half the resolution is 0.005 mm; uB = 0.005 ÷ √3 ≈ 0.0029 mm.
- Type B (calibration): the caliper’s certificate states ±0.02 mm at k = 2, so ucal = 0.010 mm.
- Combined: uc = √(0.0071² + 0.0029² + 0.010²) ≈ 0.0126 mm.
- Expanded (k = 2): U ≈ 0.025 mm.
Result: 25.43 ± 0.03 mm (k = 2).
How to Report Uncertainty
Round the uncertainty to one or two significant figures, then round the measured value to the same decimal place. State the coverage factor or confidence level so readers know what the ± means. Writing “25.4302 ± 0.025 mm” shows false precision; “25.43 ± 0.03 mm” is correct.

Related Topics
See also accuracy vs precision, types of measurement errors, significant figures in measurement, and what is calibration.
Frequently Asked Questions
What is measurement uncertainty in simple terms?
It’s the range within which the true value very likely lies, written as ± after a result.
What is the difference between Type A and Type B uncertainty?
Type A comes from statistics of repeated readings; Type B comes from other information like calibration certificates or resolution.
What does k = 2 mean?
It’s a coverage factor that expands the standard uncertainty to about 95% confidence.
Is uncertainty the same as error?
No. Error is the actual difference from the true value; uncertainty estimates how big that error could be.
How do you combine uncertainties?
Square each standard uncertainty, add them, and take the square root.